Cleaning The Collector; Cleaning The Charging Contacts - Bosch BEA 070 Original Instructions Manual

Opacimeter
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 16
!
Do not clean the optical heads or measuring cham-
ber with solvents or detergents!
3. Push the cleaning brush through the tube of the
measuring chamber until it is visible on the opposite
side.
Fig. 4:
Cleaning the tube of the measuring chamber
4. Clean the tube of the measuring chamber with the
brush.
5. Clean the glass surfaces of the two optical heads
with a clean cloth (Fig. 5).
Fig. 5:
Cleaning the optical head
6. Push the two optical heads back in.
7. Re-attach the covers.
"
The measuring chamber tube and optical heads are
clean.
Robert Bosch GmbH
5.3.5

Cleaning the collector

1. Lay the BEA 070 on its back on a clean surface
(table).
2. Remove the three screws of the collector on the
underside of the device (Fig. 6).
3. Remove the collector.
Fig. 6:
Removing the collector
4. Clean the collector with a clean cloth (Fig. 7).
Carefully remove stuck-on soot particles using a
suitable tool (e.g. screwdriver), and clean once more
with a clean cloth.
Fig. 7:
Cleaning the collector
5. Re-attach the collector to the underside of the
device, and screw in place.
"
The collector is clean.
5.3.6

Cleaning the charging contacts

Clean dirty charging contacts (on the underside of the
BEA 070) with a soft cloth only.
| BEA 070 | 25
en
|
1 689 989 113
2018-06-08