Command Syntax

All PG9000 commands are ASCII strings and are summarized in Table 21. The user must
wait for PG9000 to reply before sending another command. An exception to this is the use of
any of the IEEE Std. 488.2 Common Commands via the IEEE-488 interface (these Common
Commands are shown first in 4.3.4.1 and always start with an asterisk: "*"). The common
commands only generate a reply if using the COM1 port or if the query form of the common
command is used (command followed by a "?").
4.3.2

Command Summary

COMMAND
*CLS
*ESE(?)
*ESR?
*IDN?
*OPC(?)
*OPT?
*SRE(?)
*STB?
*RST
*TST?
ABORT
AMBHx(=)
AMBPx(=)
AMBTx(=)
AMB
AMHERR
AMHLOAD(=)
AMHMS(=)
AROT(=)
ATMHEIGHT(=)
BELL(=)
BELLx(=)
CALx(=)
COMx(=)
DATE(=)
10
<CR> and <LF> with EOI asserted with <LF>
SH1, AH1, T4, L2, RL2, DC2
Parallel poll mode is not supported
IEEE-488.2 with tri-state bus drivers
To establish IEEE-488 communications the host computer must
have an IEEE-488 card and PG9000 must be correctly addressed.
PG9000's IEEE-488 address can be set locally from the front panel
using [SPECIAL], <5Remote> (see Section 3.12.5.2). An IEEE-488
interface cable must be used.

Table 21. Command Summary

Clear the status registers and all queues
Read or set the Event Status Enable Register
Read the Event Status Register
Identify the product and software version
Read or set the Operation Complete register (not applicable to the PG9000)
Read the PG9000 options installed
Read or set the Service Request Register
Read the Status Byte
Reset user settings to factory defaults
Read the system self test results
Interrupt PG9000 activity and put it in an idle state
Set or read the source for ambient humidity
Set or read the source for ambient pressure
Set or read the source for ambient temperature
Read back the current ambient conditions
Read error messages from an AMH.
Load or unload all AMH masses
Read or set the discreet AMH mass load
Set or read the current automated motorized rotation status
Set or read the ATM head height
Read or select the mass loading bell to use
Read or set a mass loading bell's information
Set or read the calibration coefficients for an internal sensor
Set or read the COMx port configuration
Set or read the current date
Page 121
4. REMOTE OPERATION