8.7 Maintenance menu
8.7.1 Syringe activation
8.7.2 Pump activation
Select Maintenance from the home screen.
Refer also to
Table 6
on page 23.
Do a check of the syringe. Make sure the syringe fills and empties correctly.
1. From the maintenance menu, push Syringe activation. Push the arrow keys to
select an option.
Option
Description
Fill
Fills the syringe with titrant solution. The process stops automatically when
the syringe is full.
Empty to bottle
Discards the contents of the syringe into the titrant bottle.
Empty to beaker Discards the contents of the syringe into the beaker. Make sure that the
tube from the outlet port of the syringe is inside the beaker.
Stop
Stops the operation.
This option only applies to instruments with pumps installed. Do a check of the pump.
Make sure the pump fills and empties correctly.
1. From the maintenance menu, push Pump activation. Push the up and down arrow
keys to select an option.
Option Description
Start
Starts the pump. The reagent is pumped through the tubes into the beaker. Make
sure that the tube from the outlet port of the pump is in the beaker.
Stop
Stops the operation.
Maintenance
41