DIAGnostic

Subsystem Syntax

DIAGnostic:CLOSe

Parameters
Comments
Example
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The DIAGnostic subsystem contains instrument-specific commands is
are not recommended for general programming. For the E1470A, the
DIAG subsystem allows you to open/close individual relays and query
individual relays.
DIAGnostic
:CLOSe <relay>{,<relay> ...}
:CLOSe? <relay>{,<relay>...}
:OPEN <relay>{,<relay>...}
:OPEN? <relay>{,<relay>...}
:RELAY?
DIAGnostic:CLOSe <relay>{,<relay>...}
E1470A. Since these are Form C relays, "closed" means the relay is "set"
(COMMON to NO).
Name
Type
<relay>
numeric
Values other than those listed in the table cause error
Invalid Values.
2022, "Invalid relay number".
To close single relays, use DIAG:CLOS abc. To close
Closing Relays.
multiple relays, use DIAG:CLOS abc,def,ghi,... etc.
The E1470A has 80 relays. Setting more than 80
80 Relays Maximum.
relay numbers causes error: -108, "Parameter not allowed".
Closing Relays
D I A G : C L O S 0 0 1
closes individual relays on the
Range of Values
001-003|011-014|021-024|031-034|041-044|
051-056|101-103|111-114|121-124|131-134|
201-203|211-214|221-224|231-234|241-244|
251-256|301-303|311-314|321-324|331-334
!Closes relay 001 (connects
!CH001 to relay 002 in bank 00)
RF Switch Command Reference 29