Error Type/Message
Temperature Below Limit or Temperature Above
Limit During Blood Lead Test
Sensor Out of Vial Too Long
Sensor Inserted Before Ready
Sensor(s) Inserted on Exit
LeadCare Ultra
Blood Lead Testing System User's Guide
®
Reason(s)
• The temperature of the
analyzer is outside of
its operational
temperature range.
• Sensor was left in the
analyzer for more than
five (5) minutes
without use. Sensor
performance may be
compromised.
• One or more sensors
were detected in the
analyzer as the
analyzer was turned
ON.
• One or more sensors
were detected in the
analyzer as the
analyzer was turned
OFF.
Fix
• Place the system in
an environment of
ambient temperature
16°C - 30°C. Allow
the analyzer to adjust
to ambient
temperature.
• Remove and discard
sensor. Insert a fresh
sensor, enter Sample
ID and add sample to
sensor without delay.
• Remove and re-insert
sensor. If the error
continues, call
Magellan Product
Support.
• Remove and discard
sensors.
7-8