Water Protection Shroud; Main Frame Shrouds - Invacare AVIVA RX Series Service Manual

Hide thumbs Also See for AVIVA RX Series:
Table of Contents

Advertisement

Invacare® AVIVA™ RX series
1. Remove right-hand shroud.
2. Loosen screws D on left-hand side.
3. Remove shroud C forwards.
Installing Shrouds
1. Install parts in reverse order. Tighten screws to 3 Nm.
2. Test all functions.

6.7.2 Water Protection Shroud

4 mm Allen key
13 mm wrench
19 mm wrench
Oblong wooden block (at least 14 x 14 x 30 cm)
1. Place wooden block underneath mobility device and
prop up mobility device.
2. Remove both power module shrouds. See 6.7.1 Power
Module Shrouds, page 25.
3. Unplug motor cable.
4. Remove drive wheel. See 6.10.6 Replacing Drive Wheel
(5–Screw Installation), page 48.
5. Remove motor-gearbox unit. See 6.9.1 Replacing
Motor-Gearbox Unit, page 38.
6. Remove motor cable B forwards from cutout.
7. Position motor cable at point C.
8. Move shroud D carefully forwards. Pay attention that
snap hooks A at top and bottom do not break.
9. Replace shroud.
10. Install parts in reverse order.
11. Test all functions.

6.7.3 Main Frame Shrouds

4 mm Allen key
26
Fig. 6-2
Fig. 6-3
1654647-A