Fluke 15B Calibration Manual page 14

Hide thumbs Also See for 15B:
Table of Contents

Advertisement

15B & 17B
Calibration Manual
Function
AC Current (40 to 200 Hz)
g
AC Current (40 to 200 Hz)
l
AC Current (40 to 200 Hz)
a
DC Current
e
DC Current
n
DC Current
b
6. When in the 4A range, display will show 4000 counts, please ignore the last digit.
Function
AC Volts
v
DC Volts
V
6
10
V Hz Max
6
Range
400.0 µA
4000 µA
40.00 mA
400.0 mA
6
4.000 A
10.00 A
400.0 µA
4000 µA
40.00 mA
400.0 mA
6
4.000 A
10.00 A
Input
Overload
Impedance
Protection
(Nominal)
1000 V
>10 MΩ <100 pF
1000 V
>10 MΩ <100 pF
Resolution
Model 15B
0.1 µA
1.5 % + 3
1 µA
0.01 mA
1.5 % + 3
0.1 mA
0.01 A
1.5 % + 3
0.01 A
0.1 µA
1.5 % + 3
1 µA
0.01 mA
1.5 % + 3
0.1 mA
0.01 A
1.5 % + 3
0.01 A
Common Mode
Rejection Ratio
>60 dB at dc,
50 or 60 Hz
>100 dB at dc,
50 or 60 Hz
Accuracy
Model 17B
1.5 % + 3
1.5 % + 3
1.5 % + 3
1.5 % + 3
1.5 % + 3
1.5 % + 3
Normal
Mode
Rejection
_
>45 dB at
50 or 60 Hz

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

17b

Table of Contents