Agilent Technologies E4406A VSA Series Programmer's Manual page 282

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Language Reference
SENSe Subsystem
Adjacent Channel Power—Sweep Time
[:SENSe]:ACP:OFFSet:LIST:SWEep:TIME
<seconds>,<seconds>,<seconds>,<seconds>,<seconds>
[:SENSe]:ACP:OFFSet:LIST:SWEep:TIME?
Selects a specific sweep time. If you increase the sweep time, you
increase the length of the time data captured and the number of points
measured. You might need to specify a specific sweep speed to
accommodate a specific condition in your transmitter. For example, you
may have a burst signal and need to measure an exact portion of the
burst.
Selecting a specific sweep time may result in a long measurement time
since the resulting number of data points my not be the optimum 2
Use [:SENSe]:ACP:SWEep:TIME to set the number of points used for
measuring the reference channel.
You can turn off (not use) specific offsets with the
SENS:ACP:OFFSet:LIST:STATe command.
Factory Preset
and *RST:
Mode
Offset A
Basic &
11.20 ms
cdmaOne
1 µs to 50 ms
Range:
Default Unit:
seconds
Remarks:
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History:
Revision A.03.00 or later, in cdmaOne revision A.04.00
282
Offset B
Offset C
11.20 ms
11.20 ms
n
.
Offset D
Offset E
11.20 ms
11.20 ms
Chapter 5