Agilent Technologies TwisTorr 305-IC X3513-64000 User Manual page 278

Table of Contents

Advertisement

8
Technical Information
Windows Protocol
The controller can answer with the following response types:
Tab. 4
TYPE
Logic
Numeric
Alphanumeric
ACK
NACK
Unknown Window
Data Type Error
Out of Range
Win Disabled
278/302
LENGTH
VALUE
1 byte
-
6 byte
-
10 byte
-
1 byte
(0x6)
1 byte
(0x15)
1 byte
(0x32)
1 byte
(0x33)
1 byte
(0x34)
1 byte
(0x35)
TwisTorr 305-IC User Manual / 87-901-058-01
DESCRIPTION
After a read instruction of a
logic window
After a read instruction of a
numeric window
After a read instruction of
an alphanumeric window
The command execution
has been successfully
completed
The command execution
has been failed
The specified window in the
command is not a valid
window
The data type specified in
the command (Logic,
Numeric or Alphanumeric)
is not accorded with the
specified Window
The value expressed during
a write command is out of
the range value of the
specified Window
The specified window is
Read Only or temporarily
disabled (for example you
can't write the Soft Start
when the Pump is running)

Advertisement

Table of Contents
loading

Table of Contents