7
Instructions for Use

Installation of ISO-K Flange

Installation of ISO-K Flange

For ISO-K flange connections, fix the two flanges with the clamps model
IC63250DCMZ or equivalent, as shown in the following figure
(a protective screen can optionally be used).
Figure 3
Use the required number of claw clamps:
Tighten the claw clamps with a torque of 22 Nm.
214/302
4 for ISO 100 flange
4 for ISO 160 flange
TwisTorr 305-IC User Manual / 87-901-058-01