4. Error Messages and Troubleshooting
Display
Air plug is disconnected.
Arm cuff is applied too loosely.
Air is leaking from the arm cuff.
Movement during measurement and the arm
cuff has not been inflated sufficiently.
The arm cuff was inflated above 299 mmHg
when inflating the arm cuff manually.
Movement during measurement.
Clothing is interfering with the arm cuff.
Device error.
Cause
Solution
Insert the air plug securely.
Refer to section 3.1.
Apply the arm cuff tighter.
Refer to section 3.1.
Replace the arm cuff with a new one.
Refer to section 5.3.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to section 3.3.
If "E2" appears repeatedly, inflate the arm cuff
manually until it is 30 to 40 mmHg above your
previous measurement result.
Refer to section 3.3.
Do not inflate the arm cuff above 299 mmHg.
Refer to section 3.3.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to section 3.3.
Remove any clothing interfering with the arm
cuff.
Refer to section 3.1.
Contact your OMRON retail outlet or
distributor.
EN
19