Page 1
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 2
Instruction Manual PN 51-320HP/rev.J Model 320HP May 2006 High Purity pH Sensor...
Page 3
Date Notes 1/01 This is the initial release of the product manual. The manual has been refor- matted to reflect the Emerson documentation style and updated to reflect any changes in the product offering. 7/02 Updated multiple drawings throughout. 8/02 Added drawing 40105544.
MODEL 320HP TABLE OF CONTENTS MODEL 320HP HIGH PURITY PH SENSOR TABLE OF CONTENTS Section Description Page DESCRIPTION AND SPECIFICATIONS................1 Features and Applications....................1 Operation ..........................1 Sensor Specifications......................1 Ordering Information ......................2 MECHANICAL INSTALLATION ..................3 General ..........................3 Unpacking ...........................3 Preparation for Installation ....................3 Installation Inspection ......................5 OPERATION ........................11 General ..........................11...
Page 5
MODEL 320HP TABLE OF CONTENTS MODEL 320HP HIGH PURITY PH SENSOR TABLE OF CONTENTS (CONT.) LIST OF FIGURES Figure No. Title Page Liquid Junction Installation..................Model 320HP Dimensional Drawing ................ Model 320HP-50/51/54 Wiring to Models 1054A/B, 2054, 2081, 1181, 1003, 1050 Model 320HP-55 Wiring to Models 54/54e, 81, 3081, 4081, and 5081 ....
MODEL 320HP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS SECTION 1.0 GENERAL DESCRIPTION AND SPECIFICATIONS • CORROSION-RESISTANT, WEATHERPROOF CONSTRUCTION permits installation in almost any environment. • UNIQUE SENSOR DESIGN provides accurate, stable pH measurement in high purity water. • ELECTRONICALLY SHIELDED GLASS ELECTRODE protects against electrostatic noise pickup.
MODEL 320HP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS 1.4 ORDERING INFORMATION Model 320HP Flow Through High-Purity pH Sensor is designed for use when conductivity is below 50 μS. The sensor includes a 316 stainless steel flow cell, a double junction Kynar reference cell with specially designed liquid junction, pH glass electrode, temperature compensator, and a weatherproof junction box contain- ing an optional preamplifier.
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION SECTION 2.0 MECHANICAL INSTALLATION 2.1 GENERAL The sensor requires some customer assembly, there- fore care should be taken that the instructions are fol- lowed carefully to ensure all parts are properly installed. 2.2 UNPACKING Inspect the shipping container.
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION MILLIMETER INCH Shipped loose customer install and wire. For electrical DWG. NO. REV. wiring see Table above. 4320HP01 FIGURE 2-2. Model 320HP Dimensional Drawing...
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION 2.4 INSTALLATION INSPECTION Prior to operating the sensor and attempting to stan- 3. Check all electrical connections for security. Make dardize the instrument, inspect the sensor installation sure all wires are properly dressed so they do not as follows: short or ground.
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION DWG. NO. REV. 4320HP04 FIGURE 2-4. Model 320HP-55 Wiring to Models 54/54e, 81, 3081, 4081, and 5081...
Page 12
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION FIGURE 2-5. Model 320HP-55 Wiring to Xmt-P-HT-10 FIGURE 2-6. Model 320HP-58 Wiring to Xmt-P-HT-10...
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION FIGURE 2-8. Model 320HP-55 Wiring to Model 1055-22-32 FIGURE 2-9. Model 320HP-58 Wiring to Model 1055-22-32...
Page 15
MODEL 320HP SECTION 2.0 MECHANICAL INSTALLATION MODEL 320HP-55 MODEL 320HP-58 FIGURE 2-10. Model 320HP Wiring to Model 1055 Pipe/Wall Mount...
MODEL 320HP SECTION 3.0 OPERATION SECTION 3.0 OPERATION NOTE 3.1 GENERAL For the following calibration steps refer to the ana- This section contains adjustment, standardization, and set- up instructions to put the sensor in operation. The sensor is lyzer manual for adjustments of standardize, span completely automatic and may require only an occasional or efficiency.
MODEL 320HP SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING Empty the electrolyte from the reference cell and flush it 4.1 GENERAL with deionized water. This section contains the scheduled preventative mainte- PIace the reference cell on a flat surface, window side nance and troubleshooting for the sensor.
MODEL 320HP SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING 4.2.3 Electrode Maintenance. 4.3 TROUBLESHOOTING The glass electrode has a limited process life dependent on 4.3.1 Without Advanced Diagnostics. process conditions. It is our recommendation that the elec- In the event of a malfunction, refer to Table 4-1 below. This is trode be checked and replaced as required to insure the intended as a guide and lists the troubles in order of proba- proper operation of the system.
MODEL 320HP SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING TABLE 4-1. Continued (Troubleshooting ) TROUBLE CAUSE REMEDY d. Span between buffers Temperature compensator Check T.C. resistance (see Section extremely short in auto open. 4.5.5. T.C. mode. Foam junction in space. Use wood junction for buffer cal. e.
Page 20
MODEL 320HP SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING 4.3.2 Troubleshooting with Advanced Diagnostics. The Model 54/54e analyzers and Model 81, 3081, 4081, and 5081 transmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Model 1054A pH/ORP analyzer to a lesser degree. Refer to the applicable Instruction Manual for a complete description of the analyzer’s fault conditions.
MODEL 320HP SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING FIGURE 4-2. Reference Electrode Check NOTE To check the cell using a laboratory reference electrode pro- ceed as follows: The Rosemount Analytical Model 213 Millivolt Test Box provides a complete electrical simulation of any 1.
MODEL 320HP SECTION 4.0 MAINTENANCE AND TROUBLESHOOTING 4.4.3. Glass Electrode Check. The glass electrode may be checked as follows: Refer to TABLE 4-3 Figure 4-3. TEMPERATURE vs RESISTANCE OF AUTO Failure to connect the glass electrode shield to the reference T.C.
MODEL 320HP SECTION 5.0 PARTS LIST SECTION 5.0 PARTS LIST 5.1 GENERAL This section provides illustrations and listings of replacement parts for the Model 320HP Sensor. The illustrations and listings are coded with item numbers to aid in locating parts. If you are looking for a part number, look at the drawing of the assembly which contains that part and follow the arrow leading from that part to the item number.
Page 25
MODEL 320HP SECTION 5.0 PARTS LIST DWG. NO. REV. 4320HP06 FIGURE 5-1. Model 320HP (Front and Side Views) Replacement Component Locator...
6.1 GENERAL. To expedite the repair and return of Send the package prepaid to: instruments, proper communication between the cus- Emerson Process Management - Liquid Division tomer and the factory is important. A return material 2400 Barranca Parkway authorization number is required. Call (949) 757-8500.
Page 28
NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. RETURN OF MATERIAL Material returned for repair, whether in or out of warranty, should be shipped prepaid to: Emerson Process Management Liquid Division 2400 Barranca Parkway...
Page 29
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE right now. http://www.raihome.com Specifications subject to change without notice. Credit Cards for U.S. Purchases Only. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 http://www.raihome.com...
Page 30
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...