Adjust Options Menu

In the Adjust Options menu it is possible to adjust the brightness of the LCD display screen as well
as selecting the appropriate language for the market the product is sold into. To scroll to the
appropriate screen use the Options Up or Options Down button.
Note: When the product leaves the factory the language should be preset, and there should
be no reason to adjust this.
necessary to manually set the language.
To enter the Adjust Options menu press Power
to turn the machine on. The adjacent screen
should now be displayed.
To access the Adjust Options Menu push the
Options Up or Options Down button, and hold
for 2 seconds. The Adjust Options Menu will
then be displayed. One of the two screens as
shown below will be displayed.
To scroll through the different screens press the
Options Up or Options Down buttons.
Language Selection
Press the Adjust button to change the selected
language.
To lock the setting in press the
Home button.
Display Brightness
To adjust the display brightness push the
Adjust button, when the desired brightness has
been selected press the Home button to lock in
the setting.
7.1.3

Testing the Conductivity Contacts

Impedance Testing
It is possible to check the integrity of the conductivity contacts through diagnostics. Firstly enter
the Machine Status Screen as described in Section 7.1.1, and when in this mode, by touching
damp clothes or fingers across the conductivity contacts, the value adjacent to the word Touch will
increase. The minimum value is zero, this indicates an open circuit. The maximum value is 255.
No change in this value when touching the contacts, or conversely a high value when not touching
the contacts, would indicate a fault in this circuit.
Resistance Test
An alternative, but slightly less reliable method of testing the conductivity contacts, is to test the
circuit with a multimeter. When using this method the power to the dryer must be switched off.
Place one probe of the multimeter on each conductivity contact.
6.4Mohms + 2%.
To exit the diagnostic mode, press the Power button.
However when replacing the Sensor Module it will be
15
The resistance should be