Information For Commissioning The Fidamat 6-G; Checking For Leaks - Siemens FIDAMAT 6-G Operating Instructions Manual

Gas analyzer for measuring total hydrocarbon concentrations
Table of Contents

Advertisement

4.2.2

Information for commissioning the FIDAMAT 6-G

Gas preparation
Observe the information under "Gas preparation" here.
See also
Gas preparation (Page 3-6)
4.2.3

Checking for leaks

Checking for leaks
After every maintenance task which has to do with the flame ionization detector (FID) or the
gas path, you must check for leaks in the device. Proceed as follows:
FID
1. Connect the gas output to a relative pressure manometer (0 to 1 hPa, resolution
2. Open the sample gas valve. For approximately one minute, open the valve until there is
3. Wait approximately 10 minutes until pressure has equalized. Then write down the
4. Wait another minute, and read off the pressure again.
When the pressure drop within one minute does not exceed 10 ± 2 hPa, the gas path is
sufficiently tight.
Note
During measurement, the gas path must have a constant temperature, including the FID.
Test conditions:
The specified pressures are relative pressures.
Test gas:
synthetic air (sample gas path must not be contaminated)
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
0.1 hPa).
an overpressure of approx. 950 hPa and then close the valve again.
pressure.
Commissioning
4.2 Commissioning preparations
4-3

Advertisement

Table of Contents
loading