8.4 Adjustment

In practice, the usage conditions during the testing process can vary from the measurement
conditions prevailing at the time of calibration of the calibrator.
For this case, the TCL-3M165E2 offers the facility to store your own adjustment values. This
is possible for your own functions of the calibrator and for your test specimens.
The number of adjustment values can be freely selected. You can equalize the calibrator
directly on-site with your standard thermometer and thus achieve a maximum of display
accuracy. It is not necessary to send it to the manufacturer.
The calibration values set at the factory are retained. You can reset the calibrator to these
values at any time.
From the configuration range of the main window ( § 5.2), or in the "Configure DUT"
window ( § 5.5.3), you can input your own adjustment values.
IMPORTANT!
Only functions and test specimens that are marked with the
All test tasks that use adjusted functions or test specimens are affected by the
adjustments.
Technical changes reserved
Maintenance and cleaning
icon can be adjusted.