Parts and Service

The tester is warranted to be free from defects in material and workmanship for 1 year, while
under normal use. Parts and repairs are warranted for 90 days. For the complete warranty
statement, refer to the T3 Tester Instruction Sheet.
To order parts, or for warranty service, contact Fluke as follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com.

Specifications

Display Accuracy
Maximum Voltage
Between any Terminal
and Earth Ground
Input Impedance
Temperature
Altitude
Relative Humidity
Battery Type and Life
Shock, Vibration
Environmental Seal
Safety
EMC Regulations
Certifications
The LED for each range turns on by 95 % of the nominal range value.
1000 V dc; 1000 V ac rms (sine wave), Overvoltage Category III; Tester
meets requirements for CAT IV 600 V, allowing use on outdoor/direct burial
wiring or utility-side measurements.
~750 kΩ
Operating: -10 °C to +50 °C (14 °F to 122 °F) Storage: -30 °C to +60 °C
(-22 °F to +140 °F)
Operating: 3000 m (9843 ft); Storage: 10,000 m (32808 ft)
0 °C to 30 °C (32 °F to 86 °F): 90 %; 30 °C to 40 °C (86 °F to 104 °F): 75 %;
40 °C to 50 °C
(104 °F to 122 °F): 45 %
AA (2); 250 hours with NEDA 15F or IEC R6
1 m drop at 15 °C to 35 °C (59 °F to 95 °F). Sinusoidal vibration per MIL-
PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g maximum)
IP 52 per IEC 529, no vacuum applied
This tester complies with IEC 1010-1 to 1000V OVERVOLTAGE Category III,
Pollution Degree 2, and with IEC 664-1 to 600 V OVERVOLTAGE Category
IV, Pollution Degree 2.*
*OVERVOLTAGE (Installation) Categories refer to the level of Impulse
Withstand Voltage protection provided at the specified Pollution Degree.
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed
installations. Examples include switchgear and polyphase motors.
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the
installation. Examples include electricity meter and primary over-current
protection equipment.
EN61326
,
,
Listed 950Z
VDE (Pending),
N10140,
Tester
Parts and Service