Step-by-Step Test Procedure

Step 1
Insert Test Strip.
Insert a test strip, contact bars end first
and facing up, into the test port. Push it in
until it will go no further. The meter will
turn on and the display check will appear
briefly. Then the code number will appear,
followed by the
Check to make sure the code number on
the meter display matches the code number on the test strip vial.
Always confirm that the unit of measure is mg/dL when you test
your blood glucose level.
m
WARNING: If mg/dL does not appear with the
LifeScan Customer Service at 1 800 227-8862. Use of the
wrong unit of measure may cause you to misinterpret your
blood glucose level, and may lead to incorrect treatment.
Be sure the meter and test strip codes match. If they do not, code the
meter correctly. (See pages 10
m
CAUTION: Matching the code on the meter and the code on the test
strip vial is essential to obtain accurate results. Each time you test, check
to make sure the code numbers match.
28
symbol with mg/dL.
12.)
Insert This End
symbol, call