Tagging

2.1
Commissioning tag
The transmitter is supplied with a removable commissioning tag that contains
both the Device ID (the unique code that identifies a particular device in the
absence of a device tag) and a space to record the device tag (PD_TAG) (the
operational identification for the device as defined by the Piping and
Instrumentation Diagram [P&ID]).
When commissioning more than one device on a Fieldbus segment, it can be
difficult to identify which device is at a particular location. The removable tag,
provided with the transmitter, can aid in this process by linking the Device ID
to its physical location. The installer should note the physical location of the
transmitter on both the upper and lower location of the commissioning tag.
Tear off the bottom portion for each device on the segment and use it for
commissioning the segment in the control system.
Figure 2-1: Commissioning Tag
Commissioning Tag
DEVICE ID:
0011513051010001440-121698091725
DEVICE REVISION: 7.2
PHYSICAL DEVICE TAG
DEVICE ID:
0011513051010001440-121698091725
Device Barcode
DEVICE REVISION: 7.2
S / N :
PHYSICAL DEVICE TAG
A. Device revision
Quick Start Guide
Commissioning Tag
DEVICE ID:
001151AC00010001440-121698091725
DEVICE REVISION: 8.1
PHYSICAL DEVICE TAG
A
DEVICE ID:
001151AC00010001440-121698091725
Device Barcode
DEVICE REVISION: 8.1
S / N :
PHYSICAL DEVICE TAG
Quick Start Guide