Connect With Prolink Iii - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Using ProLink III with the transmitter
June 2019
Important
The user is responsible for responding to messages and notes and complying with all safety messages.
B.2 Connect with ProLink III
A connection from ProLink III to your transmitter allows you to read process data, configure the transmitter,
and perform maintenance and troubleshooting tasks.
B.2.1 Connection types supported by ProLink III
Depending on the channels that are licensed on your transmitter, you may have several connection types
available to connect from ProLink III to the transmitter. Choose the connection type appropriate to your
network and the tasks you intend to perform.
Table B-1: Connection types supported by ProLink III
Connection type
Service port
HART/Bell 202
When selecting a connection type, consider the following:
Some connections require opening the wiring compartment. These connection types should be used only
for temporary connections, and may require extra safety precautions.
Service port connections use standard connection parameters and a standard address already defined in
ProLink III.
HART/Bell 202 connections use standard HART connection parameters that are already defined in ProLink
III.
B.2.2 Make a service port connection from ProLink III to the
transmitter
CAUTION
If the transmitter is in a hazardous area, do not open the wiring compartment while the transmitter is
powered up. Opening the wiring compartment while the transmitter is powered up could cause an
explosion. To connect to the transmitter in a hazardous environment, use a connection method that does
not require opening the wiring compartment.
Prerequisites
Ensure the transmitter service port is enabled.
Obtain a USB type A to type A cable.
Important
The USB cable should be no greater than 1 meter in length.
276
Port or channel
USB A-type
Channel A (mA Output 1)
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs
Configuration and Use Manual
MMI-20039472
Terminals
N/A
1 and 2