Read section 5 entitled bath use before placing the bath in
service. Incorrect handling can damage the bath and void the

warranty.

This section gives a brief summary of the steps required to set up and operate
the 7108 bath. This should be used as a general overview and reference and not
as a substitute for the remainder of the manual. Please read Section4 through6
carefully before operating the bath.

3.1 Unpacking

Unpack the bath carefully and inspect it for any damage that may have occurred
during shipment. If there is shipping damage, notify the carrier immediately.
Verify that all components are present:
7108 Bath
Controller Probe
Access Hole Cover
Manual
If you are missing any item, please contact an Authorized Service Center (see
Section 1.4).

3.2 Set Up

Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, installing the probe and connecting the power. Allow six
inches of space on both sides of the bath for proper airflow. Consult Section4 for
detailed instructions for proper installation of the bath. Be sure to place the bath
in a safe, clean and level location.
Fill the bath tank with an appropriate liquid. For operation at moderate bath
temperatures, clean distilled water works well. Carefully pour the fluid into the
bath tank through the large rectangular access hole above the tank avoiding
spilling any fluid. The fluid must not exceed a height of 1/2 inch below the bath
lid.
The control probe must be inserted through the lid into the bath and plugged into
the socket at the back of the bath.