Fujitsu CELSIUS H720 Operating Manual page 22

Hide thumbs Also See for CELSIUS H720:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 17
6 - English
Table 3: Guidelines and manufacturer's declaration – Electromagnetic
The CELSIUS H720 is intended for operation in the electromagnetic environment specified below.
The customer or user of the CELSIUS H720 should ensure that it is used in such an environment.
Interference
IEC 60601
immunity test
Test level
Conducted RF
3 V
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
16
Downloaded from LpManual.com Manuals
Conformance
level
3 V with
eff
docking unit
and/or
peripheral
cabling, but
without
analogue
modem
3 V/m
80...1000 MHz
3 V/m
1000... 2,500
MHz
immunity
Electromagnetic environment - guidelines
Portable and mobile RF communications
equipment should be used no closer to any part
of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
a
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
CELSIUS H720
80 MHz to 800 MHz
800 MHz to 2.5 GHz
c
b
Fujitsu

Hide quick links:

Advertisement

Table of Contents
loading