DR 5000 Pour-Thru Cell Kit Installation
1-cm and 5-cm cell assembly
6
1. Arrange the multi-cell adapter so the appropriate cell holder for the Pour-Thru cell
points towards the front of the instrument. Make sure the imprint of the cell type is
visible. Do not screw down the multi-cell holder.
2. Place the Pour-Thru cell housing on top of the multi-cell holder in the DR 5000. Make
sure the locking screw holes are aligned. Secure the assembly with the two locking
screws.
3. Clean the Pour-Thru cell with a lint-free cloth. Insert the Pour-Thru cell in the cell
holder so that the cell windows coincide with those of the cell holder.
4. Route the tubing as shown in
smooth curves with no kinks or sharp bends.
5. Assemble the stand and connect the tubing, noting flow direction arrows molded into
the housing where the tubes emerge. The inflow tube is connected to the glass funnel;
the outflow tube is connected to the overflow tube assembly.
6. Connect the drain tube (black rubber tube) to the bottom fitting of the overflow tube
assembly. Connect the other end of the drain tube to an appropriate waste vessel.
The tubes can be shortened if necessary.
Figure
6. Trim the tube length as needed to achieve