EC/TDS
Measurements
60
Operational Guide
The RANGE key will change measurement from conductivity to TDS to Salinity.
Conductivity Measurements
Verify if the probe has been calibrated. Suspend the probe into the solution
to be tested. The sleeve holes must be completely submersed. Tap the probe
repeatedly to remove any air bubbles that may be trapped inside the sleeve.
To toggle between information displayed on the 3rd LCD line, use the ARROW
keys.
Temperature-compensated and absolute conductivity are available for
measurement. This is selected in the SETUP configuration.
Automatic (ATC): The conductivity probe has a built-in temperature sensor;
the temperature value is used to automatically compensate the EC/TDS
reading. When this option is selected, "ATC" tag is displayed. A temperature
coefficient for the sample must also be set.
Connect the conductivity probe to the instrument
and wait until probe parameters are loaded. The
following message is then displayed on the LCD:
"Probe Connected"
The conductivity value will be displayed on the 1st
LCD line, the temperature on the 2nd LCD line and
calibration or range specific information on the 3rd
LCD line.
If the reading is out of range when the range is
automatic, the full-scale value (200.0 mS/cm for ATC
mode or 500.0 mS/cm for absolute conductivity) will
be displayed blinking.