Parameter
Description
EC RANGE
If AUTO is used,
the edge™ will
automatically
find the correct
conductivity
range and
unit. If a fixed
value is used,
measurements
made outside
the ranges are
considered
Out of Range,
but units will
remain fixed
throughout the
measurement.
TDS RANGE
If AUTO is used,
the edge™ will
automatically
find the
correct TDS
measurement
range and
units. If a fixed
value is used,
measurements
made outside
the ranges are
considered
Out of Range,
but units will
remain fixed
throughout the
measurement.
TDS UNIT
Select units of
measurement
for TDS
measurement.
Choices
Default
AUTO,
AUTO
29.99 μS/cm,
299.9 μS/cm,
2999 μS/cm,
29.99 mS/cm,
200.0 mS/cm,
500.0 mS/cm
AUTO,
AUTO
14.99 mg/L,
149.9 mg/L,
1499 mg/L,
14.99 g/L,
100.0 g/L,
400 g/L
mg/L or ppm
ppm
EC Meter
Basic mode
Configuration
Not available but
measurement
autoranges as
needed.
Not available but
measurement
autoranges as
needed.
Available to
select.
Operational Guide