12
5.2
Connection procedure
The pluggable terminals can be removed as needed to allow more
convenient connection. To make the electrical connection, proceed as
follows:
1. Mount the instrument as described in the previous chapter
2. Connect sensor cable 1 to terminal 1/2, and where applicable,
connect the screen
3. Connect sensor cable 2 to terminal 4/5, and where applicable,
connect the screen
4. Connect switched-off power supply to terminal 16/17
5. Connect relay 1 to terminal 10/11/12
6. Connect relay 2 to terminal 13/14/15
The electrical connection is finished.
SITRANS TCSC - Operating Instructions
MAN-100414