Sterilization procedures - STERRAD
Disassembly
Disassemble the mask according to the instructions in the User Guide.
Cleaning
Alconox
AirFit F20 cushion
1. Make a solution of Alconox by diluting with drinking quality water at 1%, ie, 10 g per liter, at 68-77°F (20-25°C), according to manufacturer's instructions.
2. Soak the component in the solution for 1 minute.
3. Whilst immersed in the solution, clean with a soft bristle brush for 1 minute. Pay particular attention to all crevices and cavities.
4. Rinse by shaking vigorously in 5 liters of drinking quality water. Repeat one more time.
Inspection / Drying
1. Inspect and if required, repeat washing until visually clean.
2. Allow the mask components to air dry out of direct sunlight.
Sterilization
STERRAD
1. Package the mask components prior to sterilization as described in the manufacturer's instructions for the STERRAD Sterilization System with STERRAD chemical indicators.
Note: The use of pouches is not recommended.
2. Sterilize the mask following the manufacturer's instructions for the STERRAD Sterilization System:
• STERRAD 100S
• STERRAD NX (Standard or Advanced cycle)
Notes:
• Drying is achieved as part of the sterilization process.
• Check STERRAD indicator strips to ensure machine has run correctly.
Inspection
Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears etc), the mask component should be discarded and replaced.
Slight discoloration of the silicone components may occur and is acceptable.
Reassembly
Reassemble the mask according to the instructions in the User Guide.
Packaging and storage
Store in a dry, dust-free environment away from direct sunlight.
Storage temperature: -4°F to 140°F (-20°C to 60°C).
1
These instructions are not intended for use with the AirTouch F20 cushion.
Failure to clean the mask component as indicated may result in inadequate disinfection and sterilisation.
2
:
1
2