Figure 5
Installation of the titration stand
1
Place a magnetic stirring rod in a titration cell. Attach the
titration cell to the magnetic stirrer.
2
Position a rubber ring on the stand rod to set the
minimum height of the electrode holder. Press the
electrode holder lock button and place the electrode
holder on the stand rod.
3.4.2.1

Electrical connection of the magnetic stirrer

4. Press the button to lock the electrode holder and place the electrode holder on the
stand rod. The electrode holder is locked upon release of the button.
5. Remove the white cap from the tip of the electrode.
Insert the electrode into the electrode holder.
Warning against system damage
Make sure that the height setting of the electrode holder protects the electrode against damage
from the stirring bar and against contact with the bottom of the titration cell.
Keep the lock button pressed whenever the electrode holder is being moved.
Note: Whenever the electrode is taken from the titration stand or the titration cell is removed from
the titration stand, always place a container beneath the electrode. Otherwise, drips from the sample
or chemicals could contaminate the work area.
1. Connect the cable of the magnetic stirrer to the
TitraLab.
The power is supplied via the TitraLab.
N O T I C E
3
Insert the electrode and adjust the height of the
electrode holder.
STIRRER
Installation
port on the back of the
15