ter can become super cooled without forming a mantle. Any movement of
the super cooled water will cause a chain reaction within the cell forming
an icy slush rather than building a mantle. If this occurs, the cell must be
melted and re-chilled before building a new mantle.
• Follow all safety guidelines listed in the user's manual.
• Only trained personnel should use calibration equipment.
1.2.2

Cautions

To avoid possible damage to the instrument, follow these guidelines.
• The 2031A Quick Stick is a precision instrument. Although it has been
designed for optimum durability and trouble free operation, it must be
handled with care. After use, the plastic outer covering is very brittle.
Keep the Quick Stick in an upright supported position until it warms up to
prevent fracture from impact.
• DO NOT drop the Quick Stick. The stem could bend and may result in re-
frigerant loss.
• The Quick Stick is a relatively heavy instrument. Use the Hart model
2031-STND or other support system to prevent accidental tipping and
damage of the Quick Stick and the TPW cell. The cell and Quick Stick
must be supported separately.
1.3

Authorized Service Centers

Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS