Table 8-5
Operating pressure
Version
MAG 1100
Ceramic
● DN 2 to 65: 40 bar
(1/12" to 2½":: 580 psi)
● DN 80: 37.5 bar
(3": 540 psi)
● DN 100: 30 bar
(4": 435 psi)
Vacuum 1 x 10
(1.5 x 10
PFA
20 bar to (290 psi)
Vacuum 0.02 bar
DN 80 to 100: CO
7 bar (101.5 psi)
Table 8-6
Mechanical load/Vibration
Version
MAG 1100
● 18 to 1000 Hz random in x, y,
z, directions for 2 hours
according to EN 60068-2-36
● Sensor: 3.17 grms
● Sensor with compact MAG
5000/ 6000 mounted
transmitter: 3.17 grms
● Sensor with compact MAG
6000 I/ 6000 I Ex de mounted
transmitter: 1.14 grms
● For compact installation with
the MAG 6000 I, transmitter
to be supported to avoid
tension on sensor part
Table 8-7
Mechanical load/Vibration
Version
MAG 1100
EMC
IP67 to EN 60529 (NEMA 4X), 1
mH
O for 30 min 2004/108/EC
2
SITRANS F M MAG 1100 sensor
Operating Instructions, 01/2018, A5E03433301-AC
bar
-6
abs
psi
)
-5
abs
(0.3 psi
)
abs
abs
pressure max.
2
MAG 1100 HT
● DN 2 to 65: 40 bar
(1/12" to 2½": 580 psi)
● DN 80: 37.5 bar
(3": 540 psi)
● DN 100: 30 bar
(4": 435 psi)
Vacuum 1 x 10
bar
-6
abs
(1.5 x 10
psi
)
-5
abs
MAG 1100 HT
● 18 to 1000 Hz random in x, y,
z, directions for 2 hours
according to EN 60068-2-36
● Sensor: 3.17 grms
MAG 1100 HT
IP67 to EN 60529 (NEMA 4X), 1
mH
O for 30 min 2004/108/EC
2
Technical data
8.3 Temperature shock
MAG 1100 F
● DN 10 ... 65: 40 bar
(3/8" to 2½":: 580 psi)
● DN 80: 25 bar
(3": 363 psi)
● DN 100: 25 bar
(4": 363 psi)
Vacuum 1 x 10
bar
-6
abs
(1.5 x 10
psi
)
-5
abs
20 bar to (290 psi)
Vacuum 0.02 bar
(0.3 psi
abs
DN 80 to 100: CO
pressure max.
2
7 bar (101.5 psi)
MAG 1100 F
● 18 to 1000 Hz random in x, y,
z, directions for 2 hours
according to EN 60068-2-36
● Sensor: 3.17 grms
● Sensor with compact MAG
5000/ 6000 mounted
transmitter: 3.17 grms
● Sensor with compact MAG
6000 I/ 6000 I Ex de mounted
transmitter: 1.14 grms
● For compact installation with
the MAG 6000 I, transmitter
to be supported to avoid
tension on sensor part
MAG 1100 F
IP67 to EN 60529 (NEMA 4X), 1
mH2O for 30 min 2004/108/EC
)
abs
51