Installing/Mounting
4.6 Mounting MAG 1100 F
Note
In order to meet the 3A Sanitary Standard the sensor must be installed as a remote unit.
Note
To obtain optimum accuracy and hygenic conditions, it is important that the sensor and the
pipes are correctly aligned around their centerline, fittings must be perpendicular to the pipeline
and upstream pipelines must be straight without curves.
NOTICE
Damage to the liner
Turning the sensor around the adapter centre line after the clamp rings have been fastened
will damage the liner. The sensor may therefore only be turned when the clamp rings have
either been removed or completely loosened.
28
SITRANS F M MAG 1100 sensor
Operating Instructions, 01/2018, A5E03433301-AC