Siemens SITRANS F M MAG 1100 Operating Instructions Manual page 37

Electromagnetic flowmeter sensor
Hide thumbs Also See for SITRANS F M MAG 1100:
Table of Contents

Advertisement

5. Fit and connect electrode and coil cables as shown below.
Power supply
Transmitter
115 to 230 V AC
Outputs
Passive output
(External powered)
Active output
(Powered from transmitter)
Relay output
Digital input
Sensor connection
1)
SITRANS F M MAG 1100 sensor
Operating Instructions, 01/2018, A5E03433301-AC
11 to 30 V DC/11 to 24 V AC
L1
L
N
N
PE
Current output
(Powered from transmitter)
31
32
Digital output
Vx 3 to 30 V
56
max. 110 mA
57
R
PLC-Digital input
58
24 V max. 30 mA
Counter or PLC-Digital input
56
57
R
58
44
NO
Relay
24 V DC/1A
45
NC
42 V AC/2A
46
Common
77
11 to 30 V DC
Input
78
91 92 93 94 95 96 97
Reserved for communication modules
Electrode cable
81
Brown
82
Black
1)
0
Blue
83
84
Coil cable
Blue
85
Brown
86
Shield
Note:
Special cable with individual wire shields (shown as dotted lines) are only required
when using empty pipe function or long cables.
1
2
PE
0/4 - 20 mA
Load ≤ 800 Ω
Menu setup
Negative:
Positive:
R =
Pull up/down resistor
1 to 10 KΩ may be
required - depending on
Cables/Input resistance
Sensor
81
82
0
83
84
85
86
Connecting
5.2 Remote installation
37