To Set The Interface Temperature From The Chemstation - HP 5973 Hardware Manual

Mass selective detector
Hide thumbs Also See for 5973:
Table of Contents

Advertisement

2

To set the interface temperature from the ChemStation

To set the interface temperature from the ChemStation

Software changes
The software is revised periodically. If the steps in this procedure do not match
your MSD ChemStation software, refer to the manuals and online help supplied
with the software for more information.
1 Select
2 Click the
3 Verify that
Aux Channel.
4 Turn the heater on, and type the setpoint in the
temperature ramps.
5 The typical setpoint is 280°C.
The limits are 0°C and 350°C. A setpoint below ambient temperature turns off the
interface heater.
& $ 8 7 , 2 1
Never exceed the maximum temperature for your column.
6 Click
close the window.
7 To make the new settings part of the method, select
menu.
& $ 8 7 , 2 1
Make sure that the carrier gas is turned on and the column has been purged of air before
heating the GC/MSD interface or the GC oven.
42
Operating the MSD
Instrument Control
from the View menu.
button to display the
Aux
is selected under Type and
MSD
Apply
to download setpoints or click
`
`
Instrument
Edit
Aux: (6890)
is selected under
Thermal Aux #2
column. Do not set
Next °C
OK
to download setpoints and
Save
window.
from the Method

Advertisement

Table of Contents
loading

Table of Contents