Equipment
Multi-Product Calibrator
MET/CAL ® Metrology
Software
(see MET/CAL
Installation and Upgrade
Guide for minimum
hardware requirements)
Digital Multimeter
Pressure Module
(725 and 726)
Test Leads
Thermocouple Test
Lead
Personal computer
PC interface cable
(725 and 726)
Calibration cable
Standard Thermometer
Table 6. Required Equipment and Software
Minimum Specifications
DC voltage: 0 V to 30 V
Accuracy: ±0.005 %
DC current: 0 mA to 24 mA
Accuracy: ±0.005 %
Temperature: Type-J thermocouple
90 day accuracy: ±0.2 °C
Resistance accuracy: ±0.006 %
Frequency accuracy: ±0.01 %
Version 6.0 or later
DC voltage: 0 V to 24 V
Accuracy: ±0.0013 %
DC voltage: 0 mA to 24 mA
Accuracy: ±0.005 %
In working condition; for establishing
communication only
Two sets of stackable low thermal
EMF banana test leads
2 Type-J male miniplugs with ~12"
(30.5 cm) of 20-gauge Type-J
thermocouple wire
Windows ® 3.1 or later with terminal
communication software
Fluke 700SC serial interface cable
assembly (LEMO to DB-9 female)
Fluke 724/725 Calibration Cable
0.1 % accuracy
Recommended Model
Fluke Model 5520A Multi-Product
Calibrator only−no substitute
Contact Fluke for the latest version.
Fluke 8508A
Fluke 700PXX Series
2 red leads: Fluke PN 105809
2 black leads: Fluke PN 105806
Fluke Model 80CJ-M male miniplugs
(package of two)
20-gauge Type-J thermocouple wire
is available through an electronic
supply outlet
486 (or later) IBM-compatible
Fluke PN 667425
Fluke PN 1556747
Calibrators