Comparing the Control Solution Test Results

The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out-of-range results may occur in the following situations:
Situations
• When the control solution bottle was not
shaken well
• When the meter, test strip or control
solution was exposed to high or low
temperatures
• When the first 1 or 2 drops of control
solution was not discarded or the tip
of the bottle was not wiped clean
• When the meter is not functioning
properly
• When the control solution is past the
expiration date printed on the bottle
• When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months)
• When the control solution is
contaminated
If results continue to fall outside the range printed on the test
strip vial, the ReliOn™ Premier Test Strips and ReliOn™ Premier
BLU Blood Glucose Meter may not be working properly. Do not
use your system and contact Customer Service: 1-855-776-0662,
7 a.m. to 7 p.m. Central time, 7 days a week.
Do This
Repeat the control
solution test by
referring to the
NOTE
on page 27.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.