Operation
5.2

Menu overview

Menu level 1
Language
+SETTINGS
HEATER
for parameterizing the
instrument
Interval
Probe type
Contrast
press. indic.
Version
+OUTPUT TEST
error indication
for testing
connected
data cables
fiber optic 1
fiber optic 2
OF output 2
+INSPECTION:
+SERVICEMENU
SIGMATAX switches to
for maintenance tasks
service mode
and functional tests
+FUNCTIONSTEST
SIGMATAX switches
to service
mode
+ERROR STATUS
30
Menu level 2
Menu level 3
clean.samp. tube
cleaning probe
blowing probe
blowing valve
compressor
probe valve
ultrasonic
drain valve
filling error
delivery error
error reset
Description
Option to select the language for the menus
Self-adjusting pipe trace heater —
turn on if there is a risk of freezing!
Time interval (6–60 minutes) between
2 samples.
An interval > 12 minutes is only necessary for
longer analysis periods
Length of connection tube (10–30 m
(393.7–1181.1 in)
Reading angle (0–30) of the display;
temperature dependent
Pressure display on the screen; can be
switched off
Instrument type and program version
Switches on the potential-free fault indication
contact for 2 seconds
The fault indication contact is only held in
the previous position for as long as the
user activates it for testing purposes.
Blanks fiber optic cable 1 for 1 second
Blanks fiber optic cable 2 for 1 second
Cleans the sample tube (assisted by the
program)
Cleans the sampling probe (assisted by the
program)
Automatically cleans the sampling probe with
air
Automatically cleans the drain valve with air
Activates compressor for 30 seconds
Automatically open and close the
probe valve
Takes a new sample and activates
ultrasound for 30 seconds
Automatically opens and closes the drain
valve
Counter for filling errors that have occurred
Counter for delivery errors that have occurred
Reset the counters to "0"