Comments

Problem
AC voltage present on
the chassis
14.2
Comments
14.2.1

EMC Directive

Hart Scientifics' equipment has been tested to meet the European Electromagnetic
Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of
Conformity for your instrument lists the specific standards to which the unit was
tested.
14.2.2

Low Voltage Directive (Safety)

In order to comply with the European Low Voltage Directive (73/23/EEC), Hart
Scientific equipment has been designed to meet the EN 61010-1 and EN 61010-2-010
standards.
56
Possible Causes and Solutions
Use a wall plug tester to check the main power plug at the wall. Use an ohmmeter to
check the continuity between the ground prong on the PEM and the chassis. If the
resistance reading is greater than three ohms, there is a problem. Check the power cord
for continuity on the ground prongs. If the resistance is greater than one ohm, re place the
power cord.