50
Problem
Possible Causes and Solutions
The display shows
Controller problem. The error messages signify the following problems with the
an error code
controller.
Err 1 - a RAM error
Err 2 - a NVRAM error
Err 3 - a Structure error
Err 4 - an ADC setup error
Err 5 - an ADC ready error
Err 6 – a defective control sensor
Err 7 – a heater error
Initialize the system by performing the Factory Reset Sequence describe above.
Temperature can-
Incorrect High Limit parameter. The High Limit parameter may be set below
not be set above a
125°C. Check this value as described in Section 7.7, Operating Parameters.
certain point
Display is reading
The instrument was turned off at high temperatures and reenergized too quickly.
incorrectly
Turn the instrument off until the display is completely off and then reenergize.
Display flickers
This is normal operation and is more prevalent at high temperatures due to Seebeck
when the instru-
Coefficient of the thermoelectric cooling devices. Some time is required to fully dis-
ment is turned off
charge the switching power supply and to complete the power off cycle.
The instrument
The specifications for the instrument include an ambient temperature of 23°C. If the
does not reach
ambient temperature is above 23°C, the instrument may not be able to reach the
temperature
lowest specified temperature. Check the ambient temperature if the instrument does
not reach maximum temperature.

CE Comments

EMC Directive

Hart Scientific's equipment has been tested to meet the European Electromag-
netic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration
of Conformity for your instrument lists the specific standards to which the unit
was tested.

Low Voltage Directive (Safety)

In order to,comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN
61010-1) and IEC 1010-2-010 (EN 61010-2-010) standards.