User Mode Calibration Linc Tm Function; Calibrate Meter For User Mode; Set User Mode Calibration Linc - Hach MP-4 User Manual

Mp series portable meters
Table of Contents

Advertisement

Operation
Push UP/MS or MR/DOWN to change the time. The maximum time is 75
4.
seconds.
Push the CAL/MCLR key to accept the new auto shut off time.
5.
3.20 User mode calibration Linc
TM
The Linc
function allows for calibration when the meter is in the User mode and
the user does not have a user standard solution to calibrate the meter. This ensures
more accurate measurements. When the Linc function is used, the User mode is
linked to another standard solution. For example: If User and KCl are linked, a KCl
standard solution is used to calibrate the instrument.
Note: When a "Linc" is established for the User mode, the Linc applies to all measurement
modes using the User solution selection.

3.20.1 Calibrate meter for User mode

To calibrate the meter for the User mode:
Push the COND key, the MIN/SALT key, or the TDS key.
1.
Calibrate the meter using a standard solution (see
2.
Select the User mode (see
3.
Set the calibration Linc.
4.

3.20.2 Set User mode calibration Linc

The Linc function sets the calibration offset factor of a standard solution to the User
solution mode. The Linc stays intact in future calibrations until it is canceled (see
section 3.20.3 on page
Follow the steps below to set the KCl, NaCl or 442 calibration factor to the User
solution mode.
Push a measurement key to Linc (i.e., COND, RES, MIN/SALT or TDS).
1.
Select the User mode (see
2.
1
Trademark of Myron L Company
section 3.11 on page
28).
section 3.11 on page
TM1
function
section 4.4 on page
19).
19).
31).
27