Omron HEM-7113 Instruction Manual

Omron HEM-7113 Instruction Manual

Automatic blood pressure monitor

Advertisement

Quick Links

Arm cuff:
F
Instruction Manual
Automatic
Blood Pressure Monitor
HEM-7113
Model
5337886-7A
Package contents:
Introduction
Thank you for purchasing the OMRON HEM-7113 Digital Automatic Blood
Pressure Monitor.
I
The OMRON HEM-7113 is a fully automatic blood pressure monitor, operating
on the oscillometric principle. It measures your blood pressure and pulse rate
Display:
simply and quickly.
The unit also stores up to 14 measurements in memory.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
J.
Systolic blood pressure
It is mainly designed for general household use. Please read the Important
K.
Diastolic blood pressure
Safety Information in this instruction manual before using the unit.
L.
Memory symbol
M. Irregular Heartbeat symbol
Please read this instruction manual thoroughly before using the unit. For
N.
Heartbeat symbol
specific information about your own blood pressure, CONSULT YOUR
1.
Flashes during measurement
DOCTOR.
2.
If flashing after measurement
complete indicates blood pressure
Important Safety Information
out of recommended range
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or
arteriosclerosis.
Please read this section carefully before using the unit.
2. Preparation
2.1 Installing/Replacing the Batteries
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in
1.
Turn the main unit upside down.
death or serious injury.
(General Usage)
2.
Slide the battery cover in the direction
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
of the arrow while pressing the ribbed
• People with severe blood flow problems, or blood disorders, should consult a
part of the cover.
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
3.
Install or replace four "AAA" size
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
batteries so that the + (positive)
water. Consult a doctor immediately.
and - (negative) polarities match
Caution:
the polarities indicated on the battery
• Indicates a potentially hazardous situation which, if not avoided, may result in
compartment.
minor or moderate injury to the user or patient or damage to the equipment or
other property.
4.
Put the battery cover back in place.
(General Usage)
Slide the battery cover as indicated,
• Do not leave the unit unattended with infants or persons who cannot express
until it clicks into place.
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
Note: The measurement values
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in
continue to be stored in
incorrect measurement results.
memory even after the
• Make sure that the air tube is not wrapped around other parts of your body when
batteries are replaced.
taking measurements. This could result in injury when the air pressure in the air
2.2 Battery Life & Replacement
tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the
If the battery low symbol (
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
replace all four batteries at the same time.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
- When the battery low symbol (
• Do not operate the unit in a moving vehicle (car, airplane).
the unit for a short while. You should replace the batteries with new ones
• To inflate the cuff manually, refer to Chapter 3.3. If the cuff is over inflated, it can
ahead of time.
cause internal bleeding.
- When the symbol (
(Battery Usage)
replace the batteries with new ones at once. Turn the unit off before replacing
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of
the batteries.
clean water.
• Use only four" AAA" alkaline batteries with this unit. Do not use other types of
• Remove the batteries if the unit will not be used for three months or more.
batteries.
• Dispose of batteries according to applicable local regulations.
• Do not insert the batteries with their polarities incorrectly aligned.
Four new "AAA" alkaline batteries will last for approximately 300 measurements,
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
when used to take two measurements a day.
Since the supplied batteries are for monitoring use only, they may have
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together. General Precautions
a shorter life and not last for 300 measurements.
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising
3. Using the Unit
or eating.
3.1 Applying the Arm Cuff
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main
Be sure to wrap the arm cuff correctly so that you get accurate results.
unit not the tube itself.
Measurements can be taken in light clothing. However, please remove
• Do not inflate the arm cuff when it is not wrapped around your arm.
thick clothes, such as sweaters, before taking a reading.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the" Important information regarding Electro Magnetic
Note:
Compatibility (EMC)" in the Technical Data Section.
You can take a measurement on either your left or right arm. The blood
• Read and follow the" Correct Disposal of This Product" in the Technical Data
pressure can differ between the right arm and the left arm and therefore
Section when disposing of the device and any used accessories or optional
also the measured blood pressure values can be different. OMRON
parts.
recommends to always use the same arm for measurement. If the
values between the two arms differ substantially, please check with your
physician which arm to use for your measurement.
1. Overview
1.
Insert the air plug into the air jack
Main unit:
A
D
on the left side of the main unit.
2.
Apply the arm cuff to your upper arm. The air tube should be
centred on the inside of your arm and point down the inside of the
E
arm, so that the air tube runs down the inside of your forearm and
B
C
is in line with your middle finger.
A. Display
D. Battery compartment
B. START/STOP button.
E. Air jack
C. Memory button
If air starts to leak from the arm cuff,
3.
When the cuff is positioned correctly, close the fabric fastener
please replace with a new one.
firmly.
G
Refer to "6. Optional Parts".
Taking a Reading on the Right Arm
F. Arm cuff
(Medium cuff:
Apply the cuff so that the air tube is
arm circumference 22-32 cm)
at the side of your elbow.
G. Air tube
H
H. Air plug
Notes:
Note:
The optional arm cuff does not
• Be careful not to rest your arm on
include the air plug. Do not throw
the air tube, or otherwise restrict the
the air plug away in case
flow of air to the cuff.
purchasing the new one.
• The cuff should be 1 to 2 cm above
the elbow.
I
.
Four "AAA" alkaline (LR03)
3.2 Taking a Reading
batteries
• Instruction manual
1.
• Blood pressure monitor unit
Sit on a chair with your feet flat on
the floor and place your arm on a
table so that the arm cuff will be
J
at the same level as your heart.
K
Keep still and do not talk during measurement.
L
N
O
M
2.
Press the START/STOP button to turn on the unit.
P
Q
O.
Pulse display
P.
Deflation symbol
Q. Battery low symbol
Note: If your systolic or diastolic
pressure is outside the
standard range
Note: To cancel a measurement, press the START/STOP button to turn off
(above 135/85 mmHg),
the unit and to release the air in the arm cuff.
the Heartbeat symbol (
)
will blink. Refer to
3.
"3.2 Taking a Reading".
Check the measurement results.
The unit automatically stores blood
pressure and pulse rate into its memory.
Refer to "3.4 Using the Memory
Function".
Notes:
• Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure reading.
Waiting between readings allows the arteries to return to the condition prior
to taking the blood pressure measurement.
4.
Remove the arm cuff.
Important:
Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heart-
beats can influence the results of the
measurement. The irregular heartbeat
algorithm automatically determines if the
measurement is usable or needs to be
repeated. If the measurement results are affected by
irregular heartbeats but the result is valid, the result is
shown together with the irregular heartbeat symbol (
If the irregular heartbeats cause the measurement to be
invalid, no result is shown. If the Irregular heartbeat
symbol (
) is shown after you have taken a measure-
ment, repeat the measurement. If the Irregular heartbeat
) appears on the display,
symbol (
) is shown frequently, please make your doctor
aware of it.
) starts to blink, you will still be able to use
What is Irregular Heartbeat?
An irregular heartbeat is a heartbeat
) remains lit, the batteries are exhausted. You should
rhythm that varies by more than 25%
from the average heartbeat rhythm
detected while the unit is measuring
the systolic and diastolic blood
pressure. If such an irregular rhythm
is detected more than twice during
measurement, the irregular heartbeat
symbol (
) appears on the symbol
when the measurement results are
displayed.
What is Arrhythmia?
A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws
in the bio-electrical system that drives the heartbeat. Typical symptoms are
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia)
or slow (bradycardia) pulse. This can be caused by heart disease, aging,
physical predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only
be diagnosed by a doctor through a special examination. Whether the
appearance of the irregular heartbeat symbol (
arrhythmia or not can only be determined by an examination and diagnosis by
your doctor.
Warning:
If the irregular heartbeat symbol (
) is shown frequently, please make
your doctor aware of it. Conducting self-diagnosis and treatment based on
measurement results is dangerous. Be sure to follow the instructions of
your doctor.
5.
Press the START/STOP button to turn the unit off.
Note: If you forget to turn the unit off, it will automatically shut itself off after
five minutes.
3.3 Instructions for Special Conditions
If your systolic pressure is known to be more than 220 mmHg, press
and hold the START/STOP button until the arm cuff inflates 30 to
40 mmHg higher than your suspected systolic pressure.
1.
1 - 2 cm
Press the START/STOP button
to turn the unit on.
Measurement starts.
2.
When the cuff starts to inflate, press the START/STOP button
and keep it pressed until the pressure is 30 to 40 mmHg
higher than your suspected systolic pressure.
Note: You cannot inflate the cuff above 299 mmHg. (An error will be displayed
if you try to inflate the cuff above 299 mmHg.)
1 to 2 cm
3.
Release the START/STOP button when the cuff has been
inflated to the desired pressure.
The cuff starts to deflate and measurement starts.
4.
The rest of the procedure is the same as for normal
measurement. Refer to "3.2 Taking a Reading", steps 3 to 5.
Note: Do not apply more pressure than necessary.
3.4 Using the Memory Function
This unit has a memory capable of storing 14 sets of readings.
Note: If the memory is full, the monitor will delete the oldest readings.
1.
Press the Memory (M) button.
The result from the most recent measurement is displayed.
Important:
If your systolic or diastolic pressure is
outside the standard range, the heartbeat
symbol will blink when the measurement
result is displayed.
Refer to "3.2 Taking a Reading".
Systolic blood
Note:
If there are no measurements results
pressure
stored in memory, the screen to the right
Diastolic blood
is displayed.
pressure
Pulse display
2.
Press the Memory (M) button repeatedly to
cycle through the previous measurement
results.
The Memory number appears for a second
before the pulse rate is displayed.
The newest set is numbered "1".
At the end of the measurement, if an irregular heartbeat
is detected, the irregular heartbeat symbol will be
displayed with the result.
3.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
To Delete All the Values Stored in Memory
You cannot delete individual stored readings, all the
readings in the unit will be deleted.
).
1.
To delete stored readings, first press the Memory (M)
button.
Then while holding it down, press the START/STOP
button simultaneously for about 2-3 seconds. All readings
will then be deleted.
NormalHeartbeat
1
Pulse
2
Blood
2.
pressure
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
Irregular
Heartbeat
Short
Long
itself off after five minutes.
Pulse
Blood
4. Handling Errors and Problems
pressure
4.1 Error Messages
Error Display
Cause
Remedy
Carefully read and repeat the
Cuff is under inflated.
steps listed in section 3.3.
) in the results indicates
Repeat measurement. Remain
Movement during
still and do not talk during
measurement.
measurement.
Insert the air plug securely.
Air plug disconnected.
Refer to section 3.1.
Apply the arm cuff correctly.
Arm cuff not applied
correctly.
Refer to section 3.1.
Clothing is interfering
Remove any clothing
with the arm cuff.
interfering with the arm cuff.
Air is leaking from the
Replace cuff with new one.
arm cuff.
Arm cuff has been
Do not inflate the arm cuff
inflated above
above 299 mmHg.
299 mmHg.
Refer to section 3.3.
6. Optional Parts
Error Display
Cause
Remedy
Medium arm cuff
Arm circumference
22 - 32 cm
Replace all four "AAA"
Battery power is low.
batteries with new ones.
Refer to section 2.1.
Blinks or appears
continuously
7. Technical Data
Device error.
Product Description
Model
Display
Measurement Method
Note: The irregular heartbeat symbol (
) may also be displayed with error messages.
Measurement Range
4.2 Troubleshooting
Memory
Accuracy
Problem
Cause
Remedy
Inflation
Apply the arm cuff
Deflation
Arm cuff not applied
correctly.
correctly.
Power Source
Refer to section 3.1.
Battery life
The reading is
Remain still and do not
extremely low (or
Movement or talking
talk during
during measurement.
high).
Applied Part
measurement.
Protection Against
Remove any clothing
Clothing is interfering
interfering with the arm
Electric Shock
with the arm cuff.
cuff.
Operating
temperature/ Humidity
Make sure that the air
The air plug is not
Storage temperature/
jack is connected
securely inserted into
securely.
Humidity/ Air Pressure
Arm cuff pressure
the main unit.
Console Weight
Refer to section 3.1.
does not rise.
Cuff Weight
Outer Dimensions
Air is leaking from the
Replace the arm cuff
arm cuff.
with a new one.
Cuff Dimensions
Package Content
Apply the cuff correctly
so that it is firmly
Arm cuff deflates too
Notes:
The arm cuff is loose.
wrapped around the
soon.
• Subject to technical modification without prior notice.
arm.
Refer to section 3.1.
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
Inflate the cuff so that it
Cannot measure or
The arm cuff has
is 30 to 40 mmHg
readings are too low
not been inflated
above your previous
or too high.
measurement result.
sufficiently.
Refer to section 3.3.
Asia Pacific HQ
The unit loses power
The batteries are
Replace the batteries
during measurement.
empty.
with new ones.
The batteries are
Replace the batteries
Production Facility
empty.
with new ones.
Nothing happens
when you press the
The batteries have
Insert the batteries with
buttons.
been inserted
the correct (+/ -)
This OMRON product is manufactured under the strict quality system of OMRON
incorrectly.
polarity.
HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan for assembly.
Press the START/STOP button and repeat
measurement.
If the problem continues, try replacing the
Other problems.
batteries with new ones.
If this still does not solve the problem, contact your
local authorized OMRON distributor or dealer.
5. Maintenance and Storage
Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and cuff to extreme
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
• Do not carry out repairs of any kind by yourself.
If a defect occurs, consult your local authorized
OMRON distributor or dealer.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm cuff.
Calibration and Service
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected every
two years to ensure correct functioning and accuracy. Please
consult your local authorized OMRON distributor or dealer.
Storage
Unplug the air tube from the air jack.
Gently fold the air tube
into the arm cuff.
Note: Do not bend the air tube excessively.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
HEM-7113-AP_C_M03_120219.pdf
Large arm cuff
Arm circumference
32 - 42 cm
Automatic Blood Pressure Monitor
HEM-7113 (HEM-7113-AP)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
14 Measurements
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Automatic by electric pump
Automatic pressure release valve
4 "AAA" batteries 1.5V
Capacity of new alkaline batteries is approximately
300 measurements
= Type B
Internally powered ME equipment
10°C to 40°C/ 30 to 85% RH
-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa
Approximately 245 g without batteries
Approximately 120 g
Approximately 104 (w) mm × 64 (h) mm × 128.5 (I) mm
Approximately 146 mm × 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Main unit, cuff, Instruction manual, Battery set
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08, Alexandra Technopark,
Singapore 119967
www.omron-healthcare.com.sg
OMRON HEALTHCARE MANUFACTURING
VIETNAM CO., LTD.
Binh Duong Province, VIETNAM

Advertisement

loading