High Leak Alert; Insufficient Pressure Support Alarm; Flow Blocked Alarm - ResMed Flow Generator VPAP Adapt SV User Manual

Resmed flow generator user guide
Table of Contents

Advertisement

High Leak Alert

Triggered by:
high mask leak (greater than 30 L/min) for more than 20 seconds
the mask being removed while SmartStop is disabled.
If a high leak alert sounds:
adjust the mask to minimize leak
check that the air tube and sensor tube are connected properly.
Will stop when:
the mask leak has been rectified.
Note: The High Leak alert can be turned off in the Leak Alert option in the
Settings menu.

Insufficient Pressure Support Alarm

Triggered by:
air pressure at the mask has failed to reach an expected level for 3 breaths
the Air Circuit Learn function (page 21) has not been run after adding a new
component (eg, new mask, humidifier) to the system.
If an insufficient pressure support alarm sounds:
check that the pressure sensor tubing is not kinked
check that the pressure sensor tubing is properly connected
run LEARN CIRCUIT.
Will stop when:
the pressure sensor tubing is unobstructed and properly connected to the
flow generator and mask
treatment is stopped (Start/Stop key or SmartStop)
the VPAP Adapt SV is turned off.

Flow Blocked Alarm

Triggered by:
blockage in air circuit.
If a flow blocked alarm sounds:
1. The treatment will stop.
2. Turn power off.
3. Check whether there is a blockage in the air circuit.
4. Remove blockage.
5. Turn power back on.
6. If the flow blocked alarm activates repeatedly, discontinue use and return to
ResMed for servicing. If the alarm does not recur, then continue to use as
normal.
Alarms
27