Orp Calibration - Kensington KS-PH-221-DELUXE User Manual

Portable ph/orp meter
Table of Contents

Advertisement

ORP CALIBRATION

The KSPH-221-DELUXE meter allows one point calibration for ORP measurement. This procedure is not required
unless exact readout agreement with a work standard and at a specific ORP value is needed.
1. Press the MODE key until you see the ORP indicator
2. Rinse the ORP electrode with distilled water. Dip the electrode into the calibration solution. Stir the sensor gently.
3. Press CAL key, the meter shows the current measured value.
4. Press ▲ or ▼ key to set the displayed value. Press ENTER to confirm. The calibration indicator should begin to
flash.
5. Wait for the measured value to stabilize, the display shows "END". The meter returns measurement mode
automatically. Calibration is completed.
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
User Manual
KS-PH-221-DELUXE |
Portable pH/ORP Meter
Deluxe