Rs-232C Connector Loopback Test; Making A Loopback Test Connector; Testing The Port(S) - GE Corometrics 126 Service Manual

Corometrics 120 series
Table of Contents

Advertisement

RS-232C Connector Loopback Test

Making a Loopback Test Connector

Testing the Port(s)

Revision B
Self-Tests: RS-232C Connector Loopback Test
The Communications Setup screen includes a loopback test option for each of the
communications ports: J109, J110, and J111. Running the test requires inserting a
loopback test connector into each port being tested. (You can test more than one port
at a time.)
Start with an RJ-11C connector and bridge pin 2 (RXD) and pin 5 (TXD) together.
Figure 8-6
shows the pinout of a communications port while you face the rear panel
of a 120 Series Monitor.
Table 8-6
1.
Access the Communications Setup screen (see
2.
Insert a loopback test connector into each communications port being tested.
3.
Set the mode field on the Communications Setup screen to LOOPBACK.
Notice that the word OFF displays to the right of the mode.
4.
Verify that, after a few seconds, the status LOOPBACK OK displays. OK
indicates that the test has passed.
NOTE: If OFF remains displayed, the test has failed; contact your GE Service
Representative. Do not use this port to connect to any peripheral
equipment until the port or Communications Board has been serviced.
Table 8-6. J109, J110, and J111
Communications Ports Connector Pinout
J109, J110, Or J111 Pin #
1
2
3
4
120 Series Maternal/Fetal Monitor
2015590-001
lists the signal description for each pin.
Figure
Figure 8-6. RJ-11C Connector
Signal Description
8-7).
RTS
RXD
GND
GND
8-21