Service Mode Screens; Service Lock Screen - GE Corometrics 126 Service Manual

Corometrics 120 series
Table of Contents

Advertisement

Service Mode Screens

Service Lock Screen

Revision B
Setup Procedures: Service Mode Screens
The 120 Series Monitor is programmed with five service mode screens. These
screens are listed below:
Service Lock
Install Options
Error Log (Refer to
Communications Setup (Also, refer to
Diagnostic Control (Refer to
The Service Lock screen,
Figure
numbers and is used to access the remaining service screens. A password is required
to prevent unauthorized users from accessing the service mode screens.
To display the Service Lock screen:
1.
Select the
softkey from the General Setup screen.
SERVICE
2.
The Service Lock screen appears and the access code is displayed
as
0 0 0 0
.
3.
Use the Trim Knob control to set the access code to the current month and day.
For example: April 23 is
NOTE: The correct date and time must be set on the General Setup screen or you
will not gain access to the service screens.
4.
As soon as you enter the correct access code, the Install Options screen
displays.
Figure 5-14. Service Lock Screen
120 Series Maternal/Fetal Monitor
2015590-001
"Chapter 8,
Self-Tests".)
"Chapter 11, Peripheral
"Chapter 8,
Self-Tests".)
5-14, displays the CPU and DSP software version
0 4 2 3
0 3 2 8
. March 28 is
.
SERVICE LOCK
ENTER ACCESS CODE
0 0
0
0
CPU V3.52
DSP 03.07
Devices".)
EXIT
5-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Corometrics 128Corometrics 129