CM4 Four-Point Continuous Monitor
Symbol
Name
BF
Boron Trifluoride
3
F
Fluorine
2
TDMAT
Tetrakis(dimethylamino)titanium
H
S Dry
Hydrogen Sulfide Dry
2
H
Se
Hydrogen Selenide
2
DMA
Dimethylamine
NF
Nitrogen Trifluoride
3
UDMH
1,1 Dimethyl Hydrazine
CH
F
Methyl Fluoride
3
SO
Sulfur Dioxide
2
Table A-1: Recommended Use of Sample Filter (cont'd)

A.3 General Specifications

A.3.1 CM4
Physical dimensions are as follows:
Height: 9-3/16 inches (23.3 cm)
Width: 17 inches (43.2 cm)
Depth: 17-3/4 inches (45 cm)
Weight: 55 pounds (25 kg)
Mounting Table Top: Standard
Wall Mount: Optional
Rack Mount: Optional
Tubing dimensions are:
Sample lines: 3/16-inch (4.7 mm) I.D. x 1/4-inch (6.35 mm) O.D. FEP
Teflon, 300 ft. (90 m) maximum per point except for HBr LL 100 feet
maximum per point, and N
Exhaust line 3/16-inch (4.7 mm) I.D. x 1/4-inch (6.35 mm) O.D. tubing,
CM4 Technical Manual
Filter
Filter
Filter
Type A
Type B
Type C
X
X
X
X
X
X
X
X
X
X
H
and MMH with 50 feet maximum.
2
4
50 ft. (15m) maximum.
Output requirements are:
X
4-20 mA analog 100-800 ohms
X
Relay output contacts:
Maximum: 127 VAC or 24 VDC @ 2 amps
X
Minimum: 24 VAC or 5 VDC @ 100 mA
Installation Category (overvoltage category): II
X
Operating Temperature: 50 - 104 °F (10-40 °C)
X
Humidity: 20-80% RH
X
A.3.2 CM4-P
X
X
Physical dimensions are as follows:
X
Height: 9-3/16 inches (23.3 cm)
Width: 17 inches (43.2 cm)
Depth: 18-3/4 inches (45 cm)
Weight: 55 pounds (25 kg)
Mounting Table Top: Standard
Wall Mount: Optional
Rack Mount: Optional
Tubing dimensions are:
Sample lines: 3/16-inch (4.7 mm) I.D. x 1/4-inch (6.35 mm) O.D. FEP
Teflon, 300 ft. (90 m) maximum per point except for HBr LL 100 feet
maximum per point, and N
Exhaust line 3/16-inch (4.7 mm) I.D. x 1/4-inch (6.35 mm) O.D. tubing,
50 ft. (15m) maximum.
Output requirements are:
4-20 mA analog 100-800 ohms
Relay output contacts:
Maximum: 127 VAC or 24 VDC @ 2 amps
Minimum: 24 VAC or 5 VDC @ 100 mA
Installation Category (overvoltage category): II
A-3
H
and MMH with 50 feet maximum.