GE moisture.IQ User Manual page 144

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Appendix D. Modbus Register Map
Function
Parameter
Channel 6 /
Alarm B
5000 Block
Measurement
Read
134
Table 13: Modbus Register Map (cont.)
Sub Parameter
ID
ModBus
Address
State
602
Units
604
Type
606
Lower Setpoint
608
Upper Setpoint
610
Status
620
State
622
Units
624
Type
626
Lower Setpoint
628
Upper Setpoint
630
Block Read
0
Measurements
Block Read Unit
0
Codes
Type
Access Modbus
3602 32-bit Integer
RO
3604 32-bit Integer
RW
3606 32-bit Integer
RW
3608 32-bit float
RW
3610 32-bit float
RW
3620 32-bit Integer
RO
3622 32-bit Integer
RO
3624 32-bit Integer
RW
3626 32-bit Integer
RW
3628 32-bit float
RW
3630 32-bit float
RW
5000 32-bit float
5000 32-bit Integer
Qty. of
Notes
Request
Registers
Read Holding
2
0 = Disabled,
(0x04)
1 = Enabled
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others see tab
Read Holding
2
1 = Setpoint,
(0x03)
2 = In Band,
3 = Out Band
Read Holding
2
(0x03), Write
Multiple
(0x10)
Read Holding
2
(0x03), Write
Multiple
(0x10)
Read Input
2
1 = Energized
(0x04)
0 = De-energized
Read Holding
2
0 = Disabled,
(0x04)
1 = Enabled
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others see tab
Read Holding
2
1 = Setpoint,
(0x03)
2 = In Band,
3 = Out Band
Read Holding
2
(0x03), Write
Multiple
(0x10)
Read Holding
2
(0x03), Write
Multiple
(0x10)
Read Input
2-32
Returns 1-16
(0x04)
measurements
Read Holding
2-32
Returns 1-16 Unit
(0x03)
Codes
moisture.IQ User's Guide