To Correct Leaks in the Split Splitless Inlet
Agilent 7890 Series Troubleshooting
If the inlet fails a pressure decay test, check the following:
• Check the caps/plugs used in the test—make sure each is
correctly installed and tight.
• If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
• Check the tightness of the septum nut.
• Check the septum. Replace if old or damaged.
• Check the insert assembly installation.
• Check the liner and liner O-ring. See
O-ring on the split/splitless
• If you changed the gold seal, verify correct installation. See
To replace the gold seal on the split/splitless
• Make sure the inlet temperature remained constant during
the test.
If these items do not resolve the problem, contact Agilent for
service.
Checking for Leaks
To change the liner and
inlet.
inlet.
7
143