HP 81101A Reference Manual page 71

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Command
:VOLT[1]:OFFSet
Long
[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:OFFSet
Form
Set & Query
Parameter
Numeric
Parameter suffix
V with engineering prefixes.
Value coupling
Range coupling
With Amplitude, see
*RST value
0.0 mV
Specified Limits
–10 V to +10 V
Description
Use this command to program the offset voltage of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of voltage, you first have to
execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLtage subsystem.
The available voltage range is limited by the combination of:
• Specified current limits
• Actual OUTPUT impedance setting :OUTPut:IMPedance
• Actual expected load impedance setting
Example
To set the offset voltage:
:HOLD VOLT
:VOLT:OFF –800MV
High = Offset + Amplitude
Low = Offset – Amplitude
page 70
:OUTput:IMPedance:EXTernal
SCPI Instrument Command List
2
2
Enable VOLTAGE subsystem
Set OUTPUT offset to –800mV
Programming Reference
71