HP 81101A Reference Manual page 70

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Programming Reference
SCPI Instrument Command List
Command
:VOLT[1]
Long
[:SOURce]:VOLTage[1][:LEVel][:IMMediate][:AMPLitude]
Form
Set & Query
Parameter
Numeric
Parameter suffix
V with engineering prefixes.
Value coupling
Range coupling
With Offset, see
*RST value
1.00 V
100 mVpp to 10.0 Vpp (values are valid from 50 Ω into 50 Ω)
Specified limits
Description
Use this command to program the amplitude voltage of the output signal.
Note that to set the output levels in terms of voltage, you first have to
execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
• Specified current limits
• Actual output impedance setting :OUTPut:IMPedance
• Actual expected load impedance setting:
Example
To set the amplitude voltage:
:HOLD VOLT
:VOLT 5V
70
High = Offset +
Low = Offset –
page 71
:OUTput:IMPedance:EXTernal
Amplitude
2
Amplitude
2
Enable VOLTAGE subsystem
Set OUTPUT amplitude to 5 V

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents