T E S T D A T - Panasonic MIR-H163-PA Service Manual

Heated incubator mir-h163 series; mir-h263 series
Table of Contents

Advertisement

Test Data
MIR-H163
Chamber temperature control specification (15 points measuring) * reference data
450mm
MIR-H263
Chamber temperature control specification (15 points measuring) * reference data
600mm
-15-
<Condition>
Ambient Temperature: 20℃
Power Source: 1φ100V
Load: No load
<Measuring Points>
For ①②④⑤⑪⑫⑭⑮
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
H: 10% of internal height
For ⑥⑦⑨⑩
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
460mm
H: a half of internal heght and center
For ③⑧⑬
W・D・H: center of each layer
<Condition>
Ambient Temperature: 20℃
Power Source: 1φ100V
Load: No load
<Measuring Points>
For ①②④⑤⑪⑫⑭⑮
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
H: 10% of internal height
For ⑥⑦⑨⑩
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
510mm
H: a half of internal heght and center
For ③⑧⑬
W・D・H: center of each layer