Test Data
MIR-H163
Chamber temperature control specification (15 points measuring) * reference data
①
②
⑥
⑦
⑪
⑫
450mm
MIR-H263
Chamber temperature control specification (15 points measuring) * reference data
①
②
⑥
⑦
⑪
⑫
600mm
④
③
⑤
⑨
⑧
⑩
⑭
⑬
⑮
④
③
⑤
⑨
⑧
⑩
⑭
⑬
⑮
-15-
<Condition>
Ambient Temperature: 20℃
Power Source: 1φ100V
Load: No load
<Measuring Points>
For ①②④⑤⑪⑫⑭⑮
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
H: 10% of internal height
For ⑥⑦⑨⑩
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
460mm
H: a half of internal heght and center
For ③⑧⑬
W・D・H: center of each layer
<Condition>
Ambient Temperature: 20℃
Power Source: 1φ100V
Load: No load
<Measuring Points>
For ①②④⑤⑪⑫⑭⑮
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
H: 10% of internal height
For ⑥⑦⑨⑩
W: 80% of internal width
(each 10% from left and right)
D: 50mm from wall
510mm
H: a half of internal heght and center
For ③⑧⑬
W・D・H: center of each layer