To Replace Electrolyte Solution; Troubleshooting - Amprobe WT-50 User Manual

Dissolved oxygen water quality meter
Table of Contents

Advertisement

To replace electrolyte solution

Replace the electrolyte solution when bubbles are seen around the sensor
or membrane, meter appears to be less sensitive, or when an error message
indicates that measurements are inaccurate.
1. Remove the screw nut, washer, and o-ring from the filler port with a screw driver.
2. Flush out the current electrolyte solution.
3. Rinse the electrode by injecting distilled or de-ionized water into the filler
port. Gently shake the probe and flush out the water.
4. Use a syringe to refill the probe with electrolyte (KBr) solution from the
sprue on the probe. Refill slowly to ensure no bubble in the probe.
5. Reinstall the o-ring, washer and screw.

TROUBLESHOOTING

Can't power on
• Make sure you have connected the electrode to the meter, and the electrode
polarization time over 10 minutes.
• Check the membrane to see if air is trapped under the membrane.
• Make sure you press power key more than > 0.3 Second.
• Check the battery conditions and replace if necessary.
• Move batteries away for one minute and then re-install.